Related Experiment Video
Updated: Jun 25, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
The Sandwich spectrometer for ultra low-level gamma-ray spectrometry
J S Elisabeth Wieslander1, Mikael Hult, Joël Gasparro
1EC-JRC-IRMM, Institute for Reference Materials and Measurements, Retieseweg 111, B-2440 Geel, Belgium. elisabeth.wieslander@gmail.com
Abstract:
The technical details and performance of the newly developed Sandwich spectrometer for ultra low-level gamma-ray spectrometry are presented. The spectrometer, which consists of two HPGe detectors, an active muon shield and a lead/copper shield with a convenient and rapid opening mechanism, is located in an underground laboratory at a depth of 500 m water equivalent. The data is collected in list mode, which enables off-line data analysis to identify muon-induced events and possible Ge detector crosstalk due to Compton scattering. The background count-rate from 40 to 2700 keV normalised to the mass of the Ge crystals is 220 day(-1)kg(-1).