Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
[Classification of congenital ventricular defects using echocardiography for transcatheter closure]
Yu-qiong Lai1, Xin-sheng Huang
1Department of Cardiology, First People's Hospital of Foshan, Foshan 528000, China. lyuqiong@fsyyy.com
Objective:
To determine the anatomical variation and classification of ventricular septal defect (VSD) using echocardiography for percutaneous catheter closure in eligible cases.
Method:
The isolated ventricular septal defect was diagnosed with echocardiography in 240 patients , and 167 patients screened by transthoracic echocardiography were suitable for percutaneous catheter closure, but only 62 with isolated perimembranous VSD voluntarily received the procedure.
Results:
The procedure was successful in 58 patients, with a success rate of 93.5% with Amplatzer device. The diameter of VSD ranged from 2.4 to 13.9 (mean 5.3-/+2.0) mm with echocardiography, and the size of Amplatzer device ranged from 4-18 (mean 8.3-/+2.9) mm. Perimembranous ventricular septal defect was complicated by aneurysm formation in 22 patients. Residual trivial or mild shunt was seen in seven (12%) patients at 24 h and one (1.7%) patient at 3 months. Seven (12.1%) patients developed heart block, 3 (5.2%) had intermittence and transient complete heart block, and one had transient second degree atrioventricular block disappearing in 3 to 10 days, and 3 (5.2%) patients had complete right bundle branch block lasting for one month. None of the patients developed significant aortic regurgitation (P>0.05), although 22 showed a superior margin of the defect less than 3 mm from the aortic valve. The mean distance from the aortic valve was 3.7-/+2.7 (1.0 to 10.5) mm. No significant mitral and tricuspid regurgitation occurred in these patients. Four patients had unsuccessful procedures.
Conclusions:
Percutaneous closure with Amplatzer device can be carried out successfully in a majority of suitable defects screened using transthoracic echocardiography. Echocardiography can exactly demonstrate the anatomical variation and classification of ventricular septal defect in adults. Attention should be given to the misdiagnosis by echocardiography of a doubly committed defect as a perimembranous outflow defect. Heart block can be an important complication of the procedure.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Heart Failure IV: Classification and Diagnostic Evaluation
Cardiomyopathy I: Introduction and Classification
Cardiac Catheterization I: Pre-Procedure Overview