Relationship between amplitude and timing of heart sounds and endocardial acceleration
Aude Tassin1, Adonis Kobeissi, Luca Vitali
1University Hospital, Angers, France. audetassin@yahoo.fr
Objective:
To study the correlation between heart sounds and peak endocardial acceleration (PEA) amplitudes and timings, by modulation of paced atrioventricular (AV) delay in recipients of dual chamber pacemakers.
Methods:
Ten recipients of dual chamber pacemakers implanted for high-degree AV block were studied. Endocardial acceleration (EA) and phonocardiographic and electrocardiographic signals were recorded during performance of an AV delay scan in VDD and DDD modes.
Results:
First PEA (PEA I) and first heart sound (S1) changed similarly with the AV delay. A close intrapatient correlation was observed between S1 and PEA I amplitudes in all patients (P < 0.0001). The interpatient normalized PEA I to S1 amplitudes correlation was r = 0.89 (P < 0.0001) in DDD mode, and r = 0.81 (P < 0.0001) in VDD mode. The mean cycle-by-cycle PEA I to S1 delay was -4.3 +/- 22 ms and second PEA (PEA II) to second heart sound (S2) delay was -7.7 +/- 15 ms.
Conclusions:
A close correlation was observed between PEA I and S1 amplitudes and timings, and between PEA II and S2 timings. These observations support the hypothesis that PEA and heart sounds are manifestations of the same phenomena. EA might be a useful tool to monitor cardiac function.
Related Concept Videos
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V) valves at the...
Beats
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Cardiovascular System Abnormal Findings II: Auscultation
Abnormal Heart Sounds
Gallops: