Clinicopathological correlation of 3 patterns of calcification in a hydrophilic acrylic intraocular lens
Irmingard M Neuhann1, Thomas F Neuhann, Peter Szurman
1Tuebingen University Eye Hospital, Tuebingen, Germany. irmingard.neuhann@med.uni-tuebingen.de
Journal of Cataract and Refractive Surgery
|March 3, 2009
Abstract:
We present 3 cases of opacification in the Biocomfold 92S intraocular lens (IOL), with a documented increase in the opacification over time in 1 case. Histopathological analysis revealed the opacification was caused by calcification in the IOL's optic material. The pattern of calcium deposition, however, was different in each IOL. The causative mechanisms for this complication are unclear. Further research is warranted.