Principle, calibration, and application of the in situ alkali chloride monitor
Related Experiment Video
Updated: Jun 25, 2026

Physical, Chemical and Biological Characterization of Six Biochars Produced for the Remediation of Contaminated Sites
Published on: November 28, 2014
Christer Forsberg1, Markus Broström, Rainer Backman
1Vattenfall AB Nordic Heat, S-162 87 Stockholm, Sweden.
Abstract:
The extended use of biomass for heat and power production has caused increased operational problems with fouling and high-temperature corrosion in boilers. These problems are mainly related to the presence of alkali chlorides (KCl and NaCl) at high concentrations in the flue gas. The in situ alkali chloride monitor (IACM) was developed by Vattenfall Research and Development AB for measuring the alkali chloride concentration in hot flue gases (less than or approximately 650 degrees C). The measurement technique is based on molecular differential absorption spectroscopy in the UV range. Simultaneous measurement of SO(2) concentration is also possible. The measuring range is 1-50 ppm for the sum of KCl and NaCl concentrations and 4-750 ppm for SO(2). This paper describes the principle of the IACM as well as its calibration. Furthermore, an example of its application in an industrial boiler is given.
More Related Videos
Related Concept Videos
Flame Photometry: Lab
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Precipitation Titration Curve: Analysis
Gravimetry: Overview
In precipitation gravimetry, the analyte is converted into a precipitate and weighed. For example, the silver content in a sample can be estimated by precipitating and...