Assessment of cardiac output changes using a modified FloTrac/Vigileo algorithm in cardiac surgery patients
Alban Senn1, Danny Button, Andreas Zollinger
1Institute of Anaesthesiology and Intensive Care Medicine, Triemli City Hospital Zurich, Birmensdorferstrasse 497, 8063 Zurich, Switzerland.
Introduction:
The FloTrac/Vigileo (Edwards Lifesciences, Irvine, CA, USA) allows pulse pressure-derived cardiac output measurement without external calibration. Software modifications were performed in order to eliminate initially observed deficits. The aim of this study was to assess changes in cardiac output determined by the FloTrac/Vigileo system (FCO) with an initially released (FCOA) and a modified (FCOB) software version, as well as changes in cardiac output from the PiCCOplus system (PCO; Pulsion Medical Systems, Munich, Germany). Both devices were compared with cardiac output measured by intermittent thermodilution (ICO).
Methods:
Cardiac output measurements were performed in patients after elective cardiac surgery. Two sets of data (A and B) were obtained using FCOA and FCOB in 50 patients. After calibration of the PiCCOplus system, triplicate FCO and PCO values were recorded and ICO was determined in the supine position and cardiac output changes due to body positioning were recorded 15 minutes later (30 degrees head-up, 30 degrees head-down, supine). Student's t test, analysis of variance and Bland-Altman analysis were calculated.
Results:
Significant changes of FCO, PCO and ICO induced by body positioning were observed in both data sets. For set A, DeltaFCOA was significantly larger than DeltaICO induced by positioning the head down. For set B, there were no significant differences between DeltaFCOB and DeltaICO. For set A, increased limits of agreement were found for FCOA-ICO when compared with PCO-ICO. For set B, mean bias and limits of agreement were comparable for FCOB-ICO and PCO-ICO.
Conclusions:
The modification of the FloTrac/Vigileo system resulted in an improved performance in order to reliably assess cardiac output and track the related changes in patients after cardiac surgery.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be met...