Related Experiment Video
Updated: Jun 25, 2026

In Vivo Surface Electrocardiography for Adult Zebrafish
Published on: August 1, 2019
Intra- and interreader variability in QT interval measurement by tangent and threshold methods in a central
Gopi Krishna Panicker1, Dilip R Karnad, Mili Natekar
1Research Section, Quintiles ECG Services, Mumbai, India. gopi.panicker@quintiles.com
Background:
The QT interval can be measured by tangent (QT(Tan)) and threshold (QT(Thr)) methods; the better method is the one with lower reader variability.
Methods:
QT(Tan) and QT(Thr) were measured twice in 100 digital electrocardiograms (ECGs) by 8 experienced readers in a central laboratory. For QT(Thr), the end of the T wave was the point where the T wave reached the isoelectric baseline; for QT(Tan), it was the point where a line from the peak of the T wave through the steepest part of the descending limb intercepted the isoelectric baseline.
Results:
The average absolute intrareader variability ranged from 3.4 to 6.9 milliseconds for QT(Tan) and from 3.5 to 5.2 milliseconds for QT(Thr). By analysis of variance, intrareader SD of QT(Tan) was 7.0 and 7.5 milliseconds for QT(Thr); interreader SD was 13.1 milliseconds for QT(Tan) and 11.9 milliseconds for QT(Thr). QT(Tan) was shorter than QT(Thr) in 96 of the 100 ECGs, it exceeded QT(Thr) in 4 ECGs, which had prominent U waves.
Conclusions:
For trained readers in a central ECG laboratory using sophisticated on-screen tools for QT measurement in high-quality digital ECGs, between- and within-reader variability are comparable for QT(Tan) and QT(Thr). However, QT(Tan) is consistently shorter than QT(Thr) by up to 10 milliseconds.
More Related Videos
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...