Abstract:
NMR-based metabolomics is characterized by high throughput measurements of the signal intensities of complex mixtures of metabolites in biological samples by assaying, typically, bio-fluids or tissue homogenates. The ultimate goal is to obtain relevant biological information regarding the dissimilarity in patho-physiological conditions that the samples experience. For a long time now, this information has been obtained through the analysis of measured NMR signals via multivariate statistics. NMR data are quite complex and the use of such multivariate statistical methods as principal components analysis (PCA) for their analysis assumes that the data are multivariate normal with errors that are identical, independent and normally distributed (i.e. iid normal). There is a consensus that these assumptions are not always true for these data and, thus, several methods have been devised to transform the data or weight them prior to analysis by PCA. The structure of NMR measurement noise, or the extent to which violations of error homoscedasticity affect PCA results have neither been characterized nor investigated. A comprehensive characterization of measurement uncertainties in NMR based metabolomics was achieved in this work using an experiment designed to capture contributions of several sources of error to the total variance in the measurements. The noise structure was found to be heteroscedastic and highly correlated with spectral characteristics that are similar to the mean of the spectra and their standard deviation. A model was subsequently developed that potentially allows errors in NMR measurements to be accurately estimated without the need for extensive replication.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse.
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
High-Resolution Mass Spectrometry (HRMS)