Improving auscultatory blood pressure measurement with electronic and computer technology: the visual auscultation
Yuqi Wang1, Jin She, Haiyan Xiang
1Department of Biomedical Engineering, Beihang University, Beijing, China.
Background:
The auscultatory method is used as the reference standard in all prevalent protocols for validation of noninvasive blood pressure measuring devices, and a validation study is essentially based on the comparison between the device and observer measurements. Thus, the objectivity and accuracy of observer measurements are crucial to the validation result.
Methods:
To provide observers with more objective information about an auscultatory measurement so that sufficient information to make measurements with greater potential objectivity and accuracy can be available, a computerized data acquisition and analysis system has been developed. It cannot only acquire and store Korotkoff sound, cuff pressure, and oscillometric pulse signals, as well as the sphygmomanometer image, but it also can display the waveforms of the three signals and the sphygmomanometer video while playing the synchronous Korotkoff sounds. With this system, observers can make their measurements via the visual auscultation method, that is to say, by watching those waveforms, instead of the sphygmomanometer, while listening for synchronized Korotkoff sounds. The system was validated according to the International Protocol (IP).
Results:
The result showed that all the differences between system measurements by the visual auscultation method and observer measurements by the conventional auscultatory method were within 4 mm Hg.
Conclusions:
The visual auscultation method achieved a high degree of accuracy, and human observers can be replaced by the system in the validation study of blood pressure measuring devices.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement