Steam sterilization and internal count sheets: assessing the potential for cytotoxicity.
Anne D Lucas1, Nancy Chobin, Ramona Conner
1US Food and Drug Administration Center for Device and Radiological Health, Silver Spring, MD, USA.
AORN Journal
|March 10, 2009
Summary
Ink transfer from count sheets to surgical instruments during steam sterilization is not cytotoxic. However, ink residue removal is difficult, necessitating a barrier for instrument reuse.
Area of Science:
- Medical Devices
- Sterilization Processes
- Biocompatibility Testing
Background:
- Surgical instruments are sterilized using steam.
- Count sheets are used during sterilization procedures.
- Ink transfer from count sheets to instruments is a potential concern.
Purpose of the Study:
- To evaluate the safety of ink transfer from count sheets to surgical instruments during steam sterilization.
- To assess the cytotoxicity of ink, toner, and labels after steam sterilization.
- To determine methods for mitigating ink residue issues.
Main Methods:
- Stainless steel instruments were exposed to inked paper and subjected to steam sterilization.
- Instruments were extracted and tested for cytotoxicity.
- Preprinted labels were also examined.
Main Results:
- No significant cytotoxic response was observed in extracts from instruments exposed to ink, toner, or labels.
- Ink residue on instruments after steam sterilization proved difficult to remove.
- Ink residue was detrimental to the surgical instruments.
Conclusions:
- Ink transfer from count sheets to surgical instruments during steam sterilization does not pose a significant cytotoxic risk.
- Ink residue is problematic for instrument maintenance and longevity.
- Utilizing a barrier between count sheets and instruments is recommended to facilitate instrument reuse.
Related Concept Videos
Methods of Sterilization I: Physical Methods
As used in a healthcare facility, sterilization destroys all microorganisms through physical or chemical methods. The physical method includes steam, dry heat, boiling water, and radiation.
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
In healthcare, the chemical method of sterilization uses chemical sterilants to treat surgical instruments and medical supplies to help prevent the transmission of infectious pathogens to patients. Due to heat sensitivity, most medical supplies and equipment should not be exposed to high temperatures. These parts include rubber, plastic, glass, and other similar elements.
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...