Related Experiment Video
Updated: Jun 25, 2026

A Study of the Complexation of Mercury(II) with Dicysteinyl Tetrapeptides by Electrospray Ionization Mass Spectrometry
Published on: January 8, 2016
An MSFIA system for mercury speciation based on an anion-exchange membrane
A M Serra1, J M Estela, V Cerdà
1Department of Chemistry, Automation and the Environment Analytical Chemistry Group, University of the Balearic Islands, E-07122 Palma de Mallorca, Spain.
Abstract:
A new methodology for the in-line preconcentration, clean-up and speciation of mercury by use of an anion-exchange membrane is proposed. The speciation of mercury is based on retention of its tetrachloro complex onto the membrane while organic mercury flows freely through it. A multisyringe is used as a liquid driver and a cold vapour atomic fluorescence detector is employed to ensure a high sensitivity. Organic mercury is decomposed into to inorganic mercury by using a UV lamp. The carrier and reductant streams consist of 1.5% (m/v) hydrochloric acid and 2% (m/v) tin chloride, respectively. Certified reference material DORM-2 was digested with 37% hydrochloric acid and analysed directly without the need for extraction. The proposed system is more environmental friendly than the classical liquid-liquid extraction procedure. Mercury recoveries from spiked samples and the reference material were all close to 100%. An LOD of 14 and 16 ng/L was obtained for total and organic mercury, respectively, both with an RSD less than 1.3%.
Related Concept Videos
Ion-Exchange Chromatography
Ion Exchange
Extraction: Advanced Methods