Analytical method for perchlorate in water by liquid chromatography-mass spectrometry using an ion exchange column
Masahiro Kamoshita1, Koji Kosaka, Mari Asami
1Department of Water Supply Engineering, National Institute of Public Health, Wako, Saitama 351-0197, Japan.
Abstract:
A practical analytical method was developed for the routine analysis of perchlorate in environmental and drinking-water samples by liquid chromatography-electrospray ionization mass spectrometry (LC/ESI-MS) using an anion exchange column. By using (18)O-enriched perchlorate as an internal standard, the limits of quantification of perchlorate determined by tenfold of the signal-to-noise ratio and tenfold of the standard deviation were 0.1 and 0.03 microg L(-1), respectively. The perchlorate concentrations in the raw and finished water samples from seven water purification plants were determined by LC/ESI-MS. Perchlorate was detected in 12 out of 13 samples, and the perchlorate concentrations in the samples were from 0.1 to 36.1 microg L(-1).
More Related Videos
04:17Analysis of Organochlorine Pesticides in a Soil Sample by a Modified QuEChERS Approach Using Ammonium Formate
Published on: January 20, 2023
12:24Laboratory Estimation of Net Trophic Transfer Efficiencies of PCB Congeners to Lake Trout (Salvelinus namaycush) from Its Prey
Published on: August 29, 2014
Related Concept Videos
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Types of Detectors
Chromatographic Methods: Classification
Chromatographic techniques are typically named by...
Principles Of Column Chromatography
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...