Optimization and assessment of quantitative 124I imaging on a Philips Gemini dual GS PET/CT system
Rebecca A Gregory1, Claire A Hooker, Mike Partridge
1Joint Department of Physics, Institute of Cancer Research and Royal Marsden NHS Foundation Trust, Downs Road, Sutton, SM2 5PT, UK. rebecca.gregory@icr.ac.uk
Purpose:
Quantitative (124)I PET imaging is challenging as (124)I has a complex decay scheme. In this study the performance of a Philips Gemini dual GS PET/CT system was optimized and assessed for (124)I.
Methods:
The energy window giving the maximum noise equivalent count rate (NECR) and NEMA 2001-NU2 image quality were measured. The activity concentration (AC) accuracy of images calibrated using factors from (18)F and (124)I decaying source measurements were investigated.
Results:
The energy window 455-588 keV gave the maximum NECR of 9.67 kcps for 233 MBq. (124)I and (18)F image quality was comparable, although (124)I background variability was increased. The average underestimation in AC in (124)I images was 17.9 +/- 2.9% for nonuniform background and 14.7 +/- 2.9% for single scatter simulation (SSS) subtraction scatter correction. At 224 MBq the underestimation was 10.8 +/- 11.3%, which is comparable to 7.7 +/- 5.3% for (18)F, but increased with decreasing activity.
Conclusions:
The best (124)I PET quantitative accuracy was achieved for the optimized energy window, using SSS scatter correction and calibration factors from decaying (124)I source measurements. The quantitative accuracy for (124)I was comparable to that for (18)F at high activities of 224 MBq but diminishing with decreasing activity. Specific corrections for prompt gamma-photons may further improve the quantitative accuracy.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...