Revised delta34S reference values for IAEA sulfur isotope reference materials S-2 and S-3
Jacqueline L Mann1, Robert D Vocke, W Robert Kelly
1Inorganic Chemical Metrology Group, Analytical Chemistry Division, National Institute of Standards and Technology, 100 Bureau Drive, MS8391, Gaithersburg, MD 20899, USA. jmann@nist.gov
Abstract:
Revised delta(34)S reference values with associated expanded uncertainties (95% confidence interval (C.I.)) are presented for the sulfur isotope reference materials IAEA-S-2 (22.62 +/- 0.16 per thousand) and IAEA-S-3 (-32.49 +/- 0.16 per thousand). These revised values are determined using two relative-difference measurement techniques, gas source isotope ratio mass spectrometry (GIRMS) and double-spike multi-collector thermal ionization mass spectrometry (MC-TIMS). Gas analyses have traditionally been considered the most robust for relative isotopic difference measurements of sulfur. The double-spike MC-TIMS technique provides an independent method for value-assignment validation and produces revised values that are both unbiased and more precise than previous value assignments. Unbiased delta(34)S values are required to anchor the positive and negative end members of the sulfur delta (delta) scale because they are the basis for reporting both delta(34)S values and the derived mass-independent Delta(33)S and Delta(36)S values.
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Sulfur Assimilation
Structure and Nomenclature of Thiols and Sulfides
Mass Spectrometry: Isotope Effect
Preparation and Reactions of Sulfides
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...