Effective dose conversion factors in paediatric interventional cardiology
A Karambatsakidou1, B Sahlgren, B Hansson
1Department of Medical Physics, Karolinska University Hospital, Stockholm S-171 76, Sweden. angeliki.karambatsakidou@karolinska
Abstract:
Conversion factors for effective dose (CF(E) = effective dose/dose-area product (mSv (Gy cm2)(-1)) in paediatric interventional cardiology were estimated retrospectively for 249 patients using the dose-area product (DAP), irradiation geometry, exposure parameters and tissue-weighting factors (TWFs) from the International Commission on Radiological Protection (ICRP) 60. Two methods for estimating the conversion factors, which differed in the description of the irradiation geometry, were evaluated. The effective doses obtained with the two methods were almost identical. The results showed that irradiation geometry had no significant impact on the CF(E), and a single factor was defined for both diagnostic and interventional examinations. In addition, the effect of the new tissue-weighting factor for breast tissue (TWF(b)) given in ICRP 103 on the effective dose was assessed. The CF(E) was 3.7+/-0.2 mSv (Gy cm2)(-1) (neonate), 1.9+/-0.2 mSv (Gy cm2)(-1) (1 year), 1.0+/-0.1 mSv (Gy cm2)(-1) (5 years), 0.6+/-0.1 mSv (Gy cm2)(-1) (10 years) and 0.4+/-0.1 mSv (Gy cm2)(-1) (15 years). Applying these CFs to the individual DAP values of each patient yielded mean effective doses of 13.0 mSv (neonate), 8.6 mSv (1 year), 6.4 mSv (5 years), 8.6 mSv (10 years) and 12.7 mSv (15 years). The maximum estimated skin dose (15 patients) did not exceed 60 mGy. With the new ICRP value for TWF(b), increases in the CFs in the order of 10-30%, and in the effective dose of 10-20%, were indicated. The results indicated that the effective dose in paediatric interventional cardiology is of much greater concern than the skin dose. Furthermore, age-dependent CF(E) values are required so as not to underestimate the doses to very young patients.
Related Concept Videos
Drug Dosing: Infants and Children
IV Infusion to Oral Dosing: Conversion Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Pharmacokinetics in Pediatric Patients: Drug Excretion
Dose Size and Dosing Frequency: Determination Methods