Related Experiment Video
Updated: Jun 24, 2026

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Failure of apnea monitoring during bed-sharing
Dawn E Elder1, Peter Larsen, Jane M Zuccollo
1Department of Paediatrics and Child Health, School of Medicine and Health Sciences, University of Otago, P.O. Box 7343, Wellington, New Zealand. dawn.elder@otago.ac.nz
Abstract:
We report the case of an infant found dead in bed with an adult while attached to a respiratory monitor sensor. We experimentally re-created a bed-sharing situation with an infant mannequin to demonstrate that a Graseby home apnea monitor continues to sense adult breathing and movement in the absence of infant breathing when used in this sleep situation.