A Contusive Model of Unilateral Cervical Spinal Cord Injury Using the Infinite Horizon Impactor
Published on: July 24, 2012
A comparison of three cervical immobilization devices
David Hostler1, Deanna Colburn, S Robert Seitz
1Department of Emergency Medicine, Emergency Responder Human Performance Laboratory, University of Pittsburgh, Pittsburgh, Pennsylvania 15213, USA.
Prehospital Emergency Care
|March 18, 2009
Summary
The XCollar offers superior cervical stabilization compared to traditional collars, reducing movement in prehospital settings. This novel device may be a viable alternative when manual stabilization is not feasible.
Area of Science:
- Emergency Medicine
- Orthopedic Surgery
- Trauma Care
Background:
- Prehospital cervical spinal cord injuries (SCIs) are rare but severe.
- Spinal immobilization is standard practice for trauma patients by emergency medical services (EMS) to prevent SCI exacerbation.
- Current cervical collars are resource-intensive and may require manual stabilization.
Purpose of the Study:
- To compare the cervical stabilization properties of a novel rigid cervical immobilization collar (XCollar).
- To evaluate the XCollar against commonly used one-piece and two-piece rigid collars in the prehospital setting.
Main Methods:
- A prospective laboratory study involving 25 healthy adult volunteers.
- Goniometric techniques measured head and neck motion in seated and supine positions across horizontal, coronal, and sagittal planes.
- Statistical analysis included one-way ANOVA with Bonferroni correction for multiple comparisons (p ≤ 0.004).
Main Results:
- The XCollar permitted 10-15 mm of movement without manual stabilization, less than comparison collars.
- In the seated position, the XCollar allowed less than 10 mm of movement in sagittal and horizontal planes.
- The study included a diverse range of subject morphometrics.
Conclusions:
- The XCollar demonstrated superior cervical stabilization in healthy volunteers compared to traditional collars, without manual assistance.
- Maximal stabilization was achieved when used with a long spine board and immobilization device.
- The XCollar presents a potentially acceptable alternative to manual stabilization when EMS resources are limited.