Discomfort associated with underhumidified high-flow oxygen therapy in critically ill patients
Gerald Chanques1, Jean-Michel Constantin, Magali Sauter
1Intensive Care and Anesthesiology Department B (SAR B), Saint Eloi Hospital. Montpellier University Hospital, 80, Avenue Augustin Fliche, 34295 Montpellier Cedex 5, France. g-chanques@chu-montpellier.fr
Objective:
To measure (1) the discomfort in non-intubated patients under high-flow oxygen therapy (HFOT) humidified with bubble (BH) or heated humidifiers (HH), and (2) the hygrometric properties of oxygen with a BH and an HH.
Design And Setting:
This was a randomized cross-over study in critically ill patients during a 3-day period. The humidification device used at days 1 and 3 was changed for the other at day 2. (2) It was also an experimental bench study using the psychrometric method with five randomized flows (3, 6, 9, 12 and 15 l/min) and different humidification techniques.
Methods:
Discomfort, particularly dryness of the mouth and throat, was measured for two humidification conditions (BH and HH) using a 0-10 numerical rating scale (NRS) by patients requiring HFOT with a face mask at a flow >/=5 l/min, in a double-blinded condition.
Results:
(1) In this clinical study, 30 patients treated by HFOT at a median flow of 7.8 l/min (5.1-10.9) were included. The global incidence of moderate (NRS = 4-6) and severe discomfort (NRS = 7-10) was 25 and 29%, respectively. The median intensities of both mouth and throat dryness were significantly lower with the HH than with the BH [7.8 (5.0-9.4) vs. 5.0 (3.1-7.0), P = 0.001 and 5.8 (2.3-8.5) vs. 4.3 (2.0-5.0), P = 0.005, respectively]. (2) In the bench study, the mean absolute humidity measured at an ambient temperature of 26 degrees C with the HH was two times greater than with the BH (30 +/- 1 vs. 16 +/- 2 mg/l, P < 0.05) regardless of the flow rate.
Conclusions:
Compared to bubble humidifiers, the use of a heated-humidifier in patients with high-flow oxygen therapy is associated with a decrease of dryness symptoms mediated by increased humidity delivered to the patient.
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-IV
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue, improving...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...