Validity and reliability of devices that assess body temperature during indoor exercise in the heat
Matthew S Ganio1, Christopher M Brown, Douglas J Casa
1University of Connecticut, Storrs, CT 06269, USA. matthew.ganio@uconn.edu
Context:
When assessing exercise hyperthermia outdoors, the validity of certain commonly used body temperature measuring devices has been questioned. A controlled laboratory environment is generally less influenced by environmental factors (eg, ambient temperature, solar radiation, wind) than an outdoor setting. The validity of these temperature measuring devices in a controlled environment may be more acceptable.
Objective:
To assess the validity and reliability of commonly used temperature devices compared with rectal temperature in individuals exercising in a controlled, high environmental temperature indoor setting and then resting in a cool environment.
Design:
Time series study.
Setting:
Laboratory environmental chamber (temperature = 36.4 +/- 1.2 degrees C [97.5 +/- 2.16 degrees F], relative humidity = 52%) and cool laboratory (temperature = approximately 23.3 degrees C [74.0 degrees F], relative humidity = 40%).
Patients Or Other Participants:
Fifteen males and 10 females.
Intervention(S):
Rectal, gastrointestinal, forehead, oral, aural, temporal, and axillary temperatures were measured with commonly used temperature devices. Temperature was measured before and 20 minutes after entering the environmental chamber, every 30 minutes during a 90-minute treadmill walk in the heat, and every 20 minutes during a 60-minute rest in mild conditions. Device validity and reliability were assessed with various statistical measures to compare the measurements using each device with rectal temperature. A device was considered invalid if the mean bias (average difference between rectal and device temperatures) was more than +/-0.27 degrees C (+/-0.50 degrees F).
Main Outcome Measure(S):
Measured temperature from each device (mean and across time).
Results:
The following devices provided invalid estimates of rectal temperature: forehead sticker (0.29 degrees C [0.52 degrees F]), oral temperature using an inexpensive device (-1.13 degrees C [-2.03 degrees F]), temporal temperature measured according to the instruction manual (-0.87 degrees C [-1.56 degrees F]), temporal temperature using a modified technique (-0.63 degrees C [-1.13 degrees F]), oral temperature using an expensive device (-0.86 degrees C, [-1.55 degrees F]), aural temperature (-0.67 degrees C, [-1.20 degrees F]), axillary temperature using an inexpensive device (-1.25 degrees C, [-2.24 degrees F]), and axillary temperature using an expensive device (-0.94 degrees F [-1.70 degrees F]). Measurement of intestinal temperature (mean bias of -0.02 degrees C [-0.03 degrees F]) was the only device considered valid. Devices measured in succession (intestinal, forehead, temporal, and aural) showed acceptable reliability (all had a mean bias = 0.09 degrees C [0.16 degrees F] and r >or= 0.94]).
Conclusions:
Even during laboratory exercise in a controlled environment, devices used to measure forehead, temporal, oral, aural, and axillary body sites did not provide valid estimates of rectal temperature. Only intestinal temperature measurement met the criterion. Therefore, we recommend that rectal or intestinal temperature be used to assess hyperthermia in individuals exercising indoors in the heat.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...