Related Experiment Video
Updated: Jun 24, 2026

Quantitative Analysis of Dietary Vitamin A Metabolites in Murine Ocular and Non-Ocular Tissues Using High-Performance Liquid Chromatography
Published on: December 27, 2024
Isocratic rapid liquid chromatographic method for simultaneous determination of carotenoids, retinol, and tocopherols
Denis Thibeault1, Haixiang Su, Elizabeth MacNamara
1Department of Diagnostic Medicine, Sir Mortimer B. Davis Jewish General Hospital, Montreal, Quebec, Canada.
Abstract:
An improved isocratic and rapid HPLC method was developed for the measurement of carotenoids, retinol and tocopherols in human serum. Vitamins were extracted with hexane. Mobile phase consisted of a mixture acetonitrile:methylene chloride:methanol with 20 mM ammonium acetate. This method used a small bead size (3 microm) Spherisorb ODS2 column with titane frits. Diode array and fluorescence detectors were used respectively for the detection of carotenoids and retinol/tocopherols. Chromatographic separation was complete in 13 min for beta-cryptoxanthin, cis-trans-lycopene, alpha-carotene, beta-carotene, cis-beta-carotene, retinol, delta-tocopherol, gamma-tocopherol and alpha-tocopherol. Echinenone and tocol were employed as internal standards for diode array and fluorescence detection. Accuracy was validated using standard reference material (SRM) 968C. Intra-assay and inter-assay precision were respectively 0.2-7.3% and 3.6-12.6%. Sensitivity was verified using the ICH recommendations and the limit of detection (LOD) obtained was sufficient for routine clinical application.