Spatial resolution and radiation dose of a 64-MDCT scanner compared with published CT urography protocols
Terri J Vrtiska1, Robert P Hartman, James M Kofler
1Department of Radiology, Mayo Clinic, 200 First St., SW, Rochester, MN 55905, USA.
Objective:
The objective of our study was to compare the spatial resolution and effective dose from 64-MDCT with several published CT urography protocols.
Materials And Methods:
A phantom containing 1-, 2-, or 4-mm cylindric channels to simulate ureters with 0.25- to 3-mm plugs to simulate ureteral filling defects or ureteral diverticula was imaged using eight helical CT urography protocols. Computed radiography (CR) was also performed. Coronal maximum-intensity-projection images were created and, with the CR image, were evaluated independently by two genitourinary radiologists. Spatial resolution was evaluated by scoring each abnormality as present, visible; or as absent, not visible. Effective dose estimates for 11 CT urography protocols, including the radiographs obtained in the CT urography protocol, were calculated using published Monte Carlo organ dose coefficients.
Results:
All ureteral abnormalities detected on CR were detected on the highest-spatial-resolution reconstruction using the evaluated 64-MDCT system. The smallest filling defect identified by both was 0.25 mm. Three 0.25-mm filling defects were not detected using the evaluated 16-MDCT system. The 4-MDCT system protocols showed the poorest performance. The range of effective doses for the evaluated CT urography protocols was 20.1-66.3 mSv. The number of phases, anatomic coverage per phase, and scanning parameters all contributed to this variation in dose.
Conclusion:
The evaluated 64-MDCT system showed detection accuracy identical to that of CR. Limiting anatomic coverage for specific phases and combining phases can reduce dose for multiphase protocols by up to a factor of 2 relative to early (circa 2000) 4-MDCT.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT