Robert Donnino1, Jill E Jacobs, Jay V Doshi
1Department of Medicine, New York University School of Medicine, NYHHA VA Manhattan, 423 First Ave., 12 West Cardiology, New York, NY 10010, USA.
Objective:
Dual-source CT improves temporal resolution, and theoretically improves the diagnostic image quality of coronary artery examinations without requiring preexamination beta-blockade. The purpose of our study was to show the improved diagnostic image quality of dual-source CT compared with single-source CT despite the absence of preexamination beta-blockade in the dual-source CT group.
Materials And Methods:
We performed a retrospective analysis of consecutive patients who underwent coronary artery evaluation with either single-source CT or dual-source CT at our institution between February 2005 and October 2006. Examination reports were analyzed for the presence of image artifacts, and image quality was graded on a 3-point scale (no, mild, or severe artifact). Type of artifact (motion, calcium, quantum mottle) was also noted.
Results:
Examinations (339 single-source CT and 126 dual-source CT) of 465 patients were analyzed. Artifact was reported in 39.8% of examinations using single-source CT and in 29.4% of examinations using dual-source CT (p < 0.05). The number of examinations with motion artifact was significantly higher with single-source CT than with dual-source CT (15.9% vs 4.8%; p < 0.001) despite significantly higher heart rates in the dual-source CT group (59.4 +/- 8.4 vs 68.6 +/- 14.6 beats per minute; p < 0.001). No patients in the dual-source CT group received preexamination beta-blockade compared with 81% of patients in the single-source CT group. The presence of severe (nondiagnostic) calcium artifact was also significantly reduced in the dual-source CT group (13.0% vs 3.2%; p < 0.001).
Conclusion:
Dual-source CT provides significantly better diagnostic image quality than single-source CT despite higher heart rates in the dual-source CT group. These findings support the use of dual-source CT for coronary artery imaging without the need for preexamination beta-blockade.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...