Rapid Molecular Detection and Differentiation of Influenza Viruses A and B
Published on: January 30, 2017
Low sensitivity of rapid diagnostic test for influenza
Timothy M Uyeki1, Ramakrishna Prasad, Charles Vukotich
1Influenza Division, Centers for Disease Control and Prevention, Atlanta, Georgia, USA. tuyeki@cdc.gov
Abstract:
The QuickVue Influenza A+B Test (Quidel) was used to test nasal swab specimens obtained from persons with influenza-like illness in 3 different populations. Compared with reverse-transcriptase polymerase chain reaction, the test sensitivity was low for all populations (median, 27%; range, 19%-32%), whereas the specificity was high (median, 97%; range, 96%-99.6%).