Optimum design parameters for a therapist-constructed positive-expiratory-pressure therapy bottle device
Régis Gemerasca Mestriner1, Rafael Oliveira Fernandes, Luís Carlos Steffen
1Faculdade de Enfermagem, Nutrição e Fisioterapia, Pontifícia Universidade Católica do Rio Grande do Sul, Av Ipiranga, 6681, Prédio 12-8 degrees Andar, Partenon, Porto Alegre, Rio Grande do Sul, Brasil.
Background:
Positive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to the airway and determine whether the PEP bottle acts as a threshold resistor or a fixed-orifice resistor.
Methods:
We designed a model composed of a bottle partially filled with water, a compressed air source, a pneumotachometer, and a manometer, to evaluate the effects of various tubing diameters (range 2-25 mm inner diameter) and lengths (range 20-80 cm long). In the first set of experiments, the PEP bottle had an open top, so there was no pressure other than the atmospheric pressure against the air escaping from the immersed tubing. The distal tip of the tube was 10 cm below the surface of the water (ie, a water-column pressure of 10 cm H(2)O), and we tested flows of 1, 5, 10, 15, 20, and 25 L/min. In the second set of experiments we tested a PEP bottle, the top of which was closed except for an air-escape orifice (4, 6, 8, 9, or 10 mm).
Results:
With tubing of 2-6 mm inner diameter, the length of the tubing and the flow significantly affected the PEP pressure (ie, the system was not a threshold resistor). With tubing > or = 8 mm inner diameter there were no significant PEP-pressure differences with any of the tubing lengths or flows tested, which indicates a threshold-resistor system. The 4-mm and 6-mm air-escape orifices significantly increased the PEP pressure, whereas the 8 mm air-escape orifice did not increase the PEP pressure.
Conclusions:
To obtain a threshold-resistor PEP bottle system (ie, the PEP pressure is generated only by the water-column pressure), the tubing must be > or = 8 mm inner diameter, and the air-escape orifice must be > or = 8 mm.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Chest Physiotherapy
Purpose
CPT is primarily used for patients with excessive bronchial secretions who have difficulty clearing...