Rapid 3-D imaging of contrast flow: application in a perfused kidney phantom
N G Chen1, J B Fowlkes, P L Carson
1Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI 48109-5667, USA.
Abstract:
Previous studies indicate imaging of ultrasound contrast in 3-D is potentially superior to 2-D imaging for vascular characterization. A dual-beam, dynamic refill technique, which relies on volumetric contrast clearance and sequential imaging, was used to image a preserved porcine kidney perfused with contrast. A model was developed for the contrast profile across the renal artery to estimate fractional blood volume. This model was used along with refill curve information to measure absolute perfusion within renal cortex for a 100-cm(3) volume. Perfusion measurements from a slice within the volume were also made using a modified interval imaging technique. The measured perfusion using the dual-beam technique was consistent with the perfusion measured using the interval imaging technique (dual-beam values were 1.06 +/- 0.04 x corresponding interval imaging values). These experiments suggest that ultrasound contrast perfusion measurements are independent of the volume of contrast eliminated before refill.