The value of dual-energy bone removal in maximum intensity projections of lower extremity computed tomography
Wieland H Sommer1, Thorsten R Johnson, Christoph R Becker
1Department of Clinical Radiology, University Hospitals-Grosshadern, Ludwig-Maximilians University, Munich, Germany. wieland.sommer@med.uni-muenchen.de
Objective:
Dual-energy computed tomography (CT) makes it possible to remove bones and intraluminal plaques from angiography datasets on the basis of spectral differentiation separating iodine from calcium. The objective of this study was to evaluate the feasibility and efficiency of this technique by comparing maximum intensity projections (MIP) created with different bone removal techniques: (a) dual-energy bone removal (DEBR); (b) purely software-based bone removal without manual corrections (SBBR - MC); and (c) manually corrected software-based bone removal (SBBR + MC). A further aim was to evaluate the dual-energy-based plaque removal tool.
Materials And Methods:
Fifty-one patients underwent dual-energy CT angiography of the lower-extremity arteries on a dual-source CT scanner. CT parameters were tube potentials, 140 and 80 kVp; exposure, 80 and 340 mAs/rot; and collimation, 14 x 1.2 mm. Bolus tracking was used in the descending aorta for timing (Ultravist 370). Bones were removed from the datasets using the 3 techniques and MIP datasets were generated. Two experienced radiologists assessed image quality ((1) correct removal of bones and preservation of vessels without artificial truncation, stenoses or occlusions of arteries; (2) minor errors with residual bone in the dataset or removal of side branches; (3) significant errors impeding diagnostic evaluation), number of vessel segmentation errors, and number of nonremoved bones. Additionally, time for MIP-generation was measured. The plaque removal tool was applied to DEBR MIPs and the outcome was rated as positive, neutral, or negative.
Results:
DEBR showed better image quality than SBBR (P < 0.05; median image quality DEBR: 1; SBBR - MC: 3; SBBR + MC: 2). Less vessel segmentation errors occurred in DEBR (P < 0.05; median DEBR: 0; SBBR - MC: 5; SBBR + MC: 1). The number of nonremoved bones was not significantly different between DEBR and SBBR + MC, but significantly higher in SBBR - MC (median DEBR: 1; SBBR - MC: 2; SBBR + MC: 0). Time for generation of MIPs was lowest for SBBR - MC (P < 0.05), but also DEBR was significantly faster than manually corrected SBBR (DEBR: 160 +/- 16 seconds; SBBR - MC: 95 +/- 12 seconds; SBBR + MC: 373 +/- 69 seconds). The plaque removal tool lead to an improvement of image quality of the MIPs and a better depiction of the residual lumen in 43%.
Conclusion:
DEBR provides significant advantages, even over manually corrected SBBR. As it works completely automatically, it can effectively help to cope with the data load of CT angiography exams. Furthermore, it enables the removal of intraluminal plaques, which provides a benefit for the estimation of the residual lumen.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
X-ray Imaging
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography