Related Experiment Video
Updated: Jun 24, 2026

Lateral Molar Approach-Driven Transoral Endoscopic Procedure for Benign Infratemporal Fossa Tumor Resection
Published on: August 15, 2025
Retrieval of a detached resectoscope sheath beak during a TURP
Chi Can Huynh1, Nazim Ali Ahmad, Henry Woo
1Department of Urology, Westmead Hospital, Westmead, Sydney, NSW, Australia.
Abstract:
The detachment of a resectoscope beak during transurethral resection is a rare and unpleasant event. Published techniques for its retrieval are not practical to the majority of urologists. The authors outline a novel approach for the retrieval of a detached continuous-flow resectoscope plastic beak that is quick, safe, and effective. The plastic beak is extracted at the time of detachment using standard continuous-flow resectoscope equipment. This has increased patient safety as no further loss of time and blood resulted from the delay associated with the more conventional approaches.