Related Experiment Video
Updated: Jun 24, 2026

Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
A stable and selective electrochemical biosensor for the liver enzyme alanine aminotransferase (ALT)
Mamun Jamal1, Oliver Worsfold, Timothy McCormac
1Centre for Research in Electroanalytical Technologies (CREATE), Department of Science, Institute of Technology Tallaght, Tallaght, Dublin 24, Ireland.
Abstract:
An electrochemical method to determine alanine aminotransferase (ALT) activity over its normal and elevated physiological range was developed based upon detection of L-glutamate at a glutamate oxidase-modified platinum electrode. Measurements were carried out in the presence of ALT co-substrates L-alanine and alpha-ketoglutarate and current response from either the oxidation of hydrogen peroxide or the re-oxidation of the mediator ferrocene carboxylic acid was employed. The enzyme electrode was tested over a 6-month period and found to retain 79% of its original activity towards ALT detection with >200 measurements performed over this time. Signals associated with interfering electroactive species (ascorbic acid and uric acid) were eliminated using background subtraction at a denatured glutamate oxidase enzyme electrode. The sensitivity of the device was found to be 0.845 nA U(-1) L ALT with t(90)=180 s, linear range 10-1000 U L(-1) and LOD of 3.29 U L(-1) using amperometry at E(app)=0.4 V vs. Ag/AgCl at 308 K (35 degrees C).

