Voltammetric determination of some anti-malarial drugs using a carbon paste electrode modified with Cu(OH)2 nano-wire
Mohammad Hossein Mashhadizadeh1, Mitra Akbarian
1Tarbiat Moallem University, Tehran, Iran. mashhadizadeh@tmu.ac.ir
Abstract:
A sensitive electroanalytical methodology for the determination of chloroquine and primaquine using differential pulse voltammetry (DPV) at a Cu(OH)2 nano-wire-modified carbon paste electrode is presented. The cyclic voltammetric and DPV pulse voltammetric techniques are compared. The effects of scan rate and pH on current were investigated and an optimal scan rate of 50 mV s(-1) and a pH 5.5, 0.1 mol L(-1) phosphate buffer solution (PBS), were used. Additions of chloroquine and primaquine using DPV show linear ranges from 0.068 to 6.88 microg mL(-1) with a detection limit of 0.01 microg mL(-1) for chloroquine and 0.58-5.89 microg mL(-1) with a detection limit of 0.25 microg mL(-1) for primaquine. The method was then successfully utilized for the determination of chloroquine and primaquine in a real sample of their tablets and a recovery of 95% was obtained without interference from tablet matrix.
Related Concept Videos
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Amperometry: Overview
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Electrodeposition
Electrodeposition can...
Potentiometry: Membrane Electrodes