Adaptive cancellation of floor vibrations in standing ballistocardiogram measurements using a seismic sensor as a
Omer T Inan1, Mozziyar Etemadi, Bernard Widrow
1Department of Electrical Engineering, Stanford University, Stanford, CA 94305, USA. omeri@stanford.edu
Abstract:
An adaptive noise canceller was used to reduce the effect of floor vibrations on ballistocardiogram (BCG) measurements from a modified electronic bathroom scale. A seismic sensor was placed next to the scale on the floor and used as the noise reference input to the noise canceller. BCG recordings were acquired from a healthy subject while another person stomped around the scale, thus causing increased floor vibrations. The noise canceller substantially eliminated the artifacts in the BCG signal due to these vibrations without distorting the morphology of the measured BCG. Additionally, recordings were obtained from another subject standing inside a parked bus while the engine was running. The artifacts due to the vibrations of the engine, and the other vehicles moving on the road next to the bus, were also effectively eliminated by the noise canceller. The system with automatic floor vibration cancellation could be used to increase BCG measurement robustness in home monitoring applications. Additionally, the noise cancellation approach may enable BCG recording in ambulances-or other transport vehicles-where noninvasive hemodynamic monitoring may otherwise not be feasible.
Related Concept Videos
Equilibrium and Balance
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...