Respiratory monitoring by inductive plethysmography in unrestrained subjects using position sensor-adjusted
Gregor Brüllmann1, Karsten Fritsch, Robert Thurnheer
1Pulmonary Division, Department of Internal Medicine, University Hospital Zurich, CH-8091 Zurich, Switzerland.
Background:
Portable respiratory inductive plethysmography (RIP) is promising for noninvasive monitoring of breathing patterns in unrestrained subjects. However, its use has been hampered by requiring recalibration after changes in body position.
Objectives:
To facilitate RIP application in unrestrained subjects, we developed a technique for adjustment of RIP calibration using position sensor feedback.
Methods:
Five healthy subjects and 12 patients with lung disease were monitored by portable RIP with sensors incorporated within a body garment. Unrestrained individuals were studied during 40-60 min while supine, sitting and upright/walking. Position was changed repeatedly every 5-10 min. Initial qualitative diagnostic calibration followed by volume scaling in absolute units during 20 breaths in different positions by flow meter provided position-specific volume-motion coefficients for RIP. These were applied during subsequent monitoring in corresponding positions according to feedback from 4 accelerometers placed at the chest and thigh. Accuracy of RIP was evaluated by face mask pneumotachography.
Results:
Position sensor feedback allowed accurate adjustment of RIP calibration during repeated position changes in subjects and patients as reflected in a minor mean difference (bias) in breath-by-breath tidal volumes estimated by RIP and flow meter of 0.02 liters (not significant) and limits of agreement (+/-2 SD) of +/-19% (2,917 comparisons). An average of 10 breaths improved precision of RIP (limits of agreement +/-14%).
Conclusions:
RIP calibration incorporating position sensor feedback greatly enhances the application of RIP as a valuable, unobtrusive tool to investigate respiratory physiology and ventilatory limitation in unrestrained healthy subjects and patients with lung disease during everyday activities including position changes.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...