Related Experiment Video
Updated: Jun 24, 2026

Simultaneous Label-Free Autofluorescence Multi-Harmonic Microscopy
Published on: August 29, 2025
Enhanced fluorescence signal in nonlinear microscopy through supplementary fiber-optic light collection
Christoph J Engelbrecht1, Werner Göbel, Fritjof Helmchen
1Department of Neurophysiology, Brain Research Institute, University of Zürich, Winterthurerstrasse 190, CH-8057 Zürich, Switzerland.
Abstract:
Nonlinear microscopy techniques crucially rely on efficient signal detection. Here, we present a ring of large-core optical fibers for epi-collection of fluorescence photons that are not transmitted through the objective and thus normally wasted. Theoretical treatments indicated that such a supplementary fiber-optic light collection system (SUFICS) can provide an up to 4-fold signal gain. In typical in vivo imaging experiments, the fiber-ring channel was brighter than the objective channel down to 800 microm depth, thus providing a gain >2. Moreover, SUFICS reduced noise levels in calcium imaging experiments by about 23%. We recommend SUFICS as a generally applicable, effective add-on to nonlinear microscopes for enhancing fluorescence signals.
More Related Videos
Related Concept Videos
Super-resolution Fluorescence Microscopy
Confocal Fluorescence Microscopy
Immunofluorescence Microscopy
The...