Image correction during large and rapid B(0) variations in an open MRI system with permanent magnets using navigator
1Shanghai Key Laboratory of Functional Magnetic Resonance Imaging and Department of Physics, East China Normal University, Shanghai, China. jqli@phy.ecnu.edu.cn
Abstract:
An open permanent magnet system with vertical B(0) field and without self-shielding can be quite susceptible to perturbations from external magnetic sources. B(0) variation in such a system located close to a subway station was measured to be greater than 0.7 microT by both MRI and a fluxgate magnetometer. This B(0) variation caused image artifacts. A navigator echo approach that monitored and compensated the view-to-view variation in magnetic resonance signal phase was developed to correct for image artifacts. Human brain imaging experiments using a multislice gradient-echo sequence demonstrated that the ghosting and blurring artifacts associated with B(0) variations were effectively removed using the navigator method.