Establishment of a Murine Pulp Exposure Model with a Novel Mouth-Gag for Pulpitis Research
Published on: October 27, 2023
A novel instrument to determine pulp proximity
David G Purton1, Nicholas P Chandler, Brian D Monteith
1david.purton@dent.otago.ac.nz
Abstract:
The aim was to determine whether extracted teeth could be used to test the Prepometer instrument, which indicates pulp proximity with green/amber/red light emitting diodes. Third molars were reduced to a plane in dentine and Prepometer readings made. Dentine was removed in 0.5 mm increments and readings made until only lights 9 or 10 (red) showed. The teeth were sectioned and the dentine thickness measured. Analysis permitted construction of a highly significant predictor-model (p < 0.01), the red/amber light boundary coinciding with a dentine thickness of 2.4 mm. The Prepometer was consistent in predicting pulp proximity but was more sensitive than specified.
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: