Objective:
The Artificial Valve Endocarditis Reduction Trial was stopped on January 21, 2000, due to a higher incidence of paraprosthetic leak in the St Jude Medical Silzone prosthesis compared with the conventional prosthesis. The Artificial Valve Endocarditis Reduction Trial investigators reported the 2-year results in 2002. This retrospective study assessed the influence on thromboembolism and paraprosthetic leak to 7 years.
Methods:
A total of 253 patients had 254 operations: 80 aortic valve replacements, 139 mitral valve replacements, and 35 multiple replacements with placement of Silzone prostheses. The mean age was 58.6 years (range 21-84 years, median age 59.8 years), and there were 126 women (49.8%) and 74 concomitant procedures (coronary artery bypass 28.9%).
Results:
Major paraprosthetic leak (repair, re-replacement, or mortality) occurred in 10 of the original procedures after 30 days (3 aortic valve replacements, 3 mitral valve replacements, 4 multiple replacements). Nine occurrences in 8 patients-5 early (
Conclusions:
Paraprosthetic leak with the St Jude Medical Silzone prosthesis was managed both during the early (