Background:
A variety of femoral stem designs have been reported to be successful in revision total hip arthroplasty without consensus as to optimal design. We evaluated the clinical and radiographic outcomes in a consecutive series of femoral revisions using a wedge-shape, tapered-stem design at medium and long-term follow-up.
Materials And Methods:
We performed a retrospective review of clinical and radiographic outcomes of twenty-eight consecutive femoral revisions arthroplasties, which were done using the Zweymuller femoral stem.
Results:
The mean follow-up was 7.4 years (range 2-15 years). No stem re-revision was necessary. All stems were judged to be stable by radiographic criteria at the most recent follow-up. The final mean Harris hip score was 90. There was no difference in Harris hip scores, implant stability, or radiological appearance (distal cortical hypertrophy or proximal stress shielding) of the implants between medium-term (mean 5.7 years) and long-term (mean 12.4 years) follow-up.
Conclusions:
We found the Zweymuller femoral stem design to be durable for revision hip arthroplasty when there is an intact metaphyseal-diaphyseal junction for adequate press-fit stability at surgery.