Evaluation of the new Vitek 2 ANC card for identification of medically relevant anaerobic bacteria
Related Experiment Video
Updated: Jun 23, 2026

Direct Microbial Identification using An Automated Microbial Identification System to Facilitate the EUCAST RAST Method Without Mass Spectrometry
Published on: May 24, 2024
Francine Mory1, Corentine Alauzet, Céline Matuszeswski
1Laboratoire de Bactériologie, Hôpital Central, 29 Avenue du Maréchal de Lattre de Tassigny, 54035 Nancy Cedex, France.
Abstract:
Of 261 anaerobic clinical isolates tested with the new Vitek 2 ANC card, 257 (98.5%) were correctly identified at the genus level. Among the 251 strains for which identification at the species level is possible with regard to the ANC database, 217 (86.5%) were correctly identified at the species level. Two strains (0.8%) were not identified, and eight were misidentified (3.1%). Of the 21 strains (8.1%) with low-level discrimination results, 14 were correctly identified at the species level by using the recommended additional tests. This system is a satisfactory new automated tool for the rapid identification of most anaerobic bacteria isolated in clinical laboratories.