Accuracy of non-invasive ICP assessment can be increased by an initial individual calibration
B Schmidt1, M Weinhold, M Czosnyka
1Department of Neurology, Medical Centre Chemnitz, Dresdner Str. 178, 09131 Chemnitz, Germany. B.Schmidt@skc.de
Objective:
In a formerly introduced mathematical model, intracranial pressure (ICP) could be non-invasively assessed using cerebral blood flow velocity (FV) and arterial blood pressure (ABP). The current study attempts to check whether the accuracy of the non-invasive ICP assessment (nICP) improves after an initial individual calibration by implanted ICP probes.
Methods:
Thirteen patients with brain lesions (35-77 years, mean: 58 +/- 13 years) were studied. FV, ABP and ICP signals were recorded at days 1, 2, 4 and 7. nICP was calculated and compared to ICP. In the first recording of each patient the (invasively assessed) ICP signal was used to calibrate the nICP calculation procedure, while the follow-up recordings were used for its validation.
Findings:
In 11 patients 22 follow-up recordings were performed. The mean deviation between ICP and the original nICP (+/- SD) was 8.3 +/- 7.9 mmHg. Using the calibrated method this deviation was reduced to 6.7 +/- 6.7 mmHg (P < 0.005).
Conclusions:
Initial individual calibration of nICP assessment method significantly improves the accuracy of nICP estimation on subsequent days. This hybrid method of ICP assessment may be used in intensive care units in patients with initially implanted ICP probes. After removal of the probes, ICP monitoring can be continued using the calibrated nICP assessment procedure.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...