The Fenwal CS3000 inline pack is acceptable for donor blood sampling
P J Dayton1, M J Randels, R G Strauss
1Elmer L. DeGowin Memorial Blood Center, University of Iowa Hospitals and Clinics, Iowa City 52242.
Journal of Clinical Apheresis
|January 1, 1991
Summary
Blood sampling packs on apheresis kits allow accurate same-day donor testing. This closed system method for complete blood counts (CBC) is reliable, even with minor contamination.
Area of Science:
- Hematology
- Transfusion Medicine
- Medical Device Technology
Background:
- Closed system apheresis kits for the Fenwal CS3000 blood cell separator offer a blood sampling pack.
- This pack aims to facilitate same-day donor testing while maintaining a closed system.
- Potential concerns include sample accuracy due to clotting, cell adherence, or dilution with priming fluid.
Purpose of the Study:
- To evaluate the accuracy of complete blood counts (CBC) using blood samples collected from an inline pack on apheresis kits.
- To compare CBC results from the inline pack with those from simultaneous venipuncture samples.
- To assess the impact of potential contaminants (e.g., saline) on CBC accuracy from the inline pack.
Main Methods:
- Paired blood samples were collected simultaneously from 17 donors using the inline pack and separate venipuncture.
- Complete blood counts (CBC) were performed on both sets of samples.
- Simulated contamination with saline was introduced to assess its effect on pack sample accuracy.
Main Results:
- CBC values obtained from the inline pack and venipuncture were nearly identical when samples were collected properly.
- Even with deliberate saline contamination, CBC values from the pack showed only a slight decrease.
- The inline sampling method maintained system closure throughout the process.
Conclusions:
- The inline blood sampling pack on closed system apheresis kits provides accurate donor blood samples for CBC analysis.
- This method effectively maintains a closed system, reducing risks of contamination and ensuring donor safety.
- The findings support the use of inline packs for reliable same-day donor testing in apheresis procedures.