Comparative evaluation of the geometrical accuracy of intravascular magnetic resonance imaging: a phantom study
Guillaume Gilbert1, Gilles Soulez, Gilles Beaudoin
1Département de Radiologie, Centre Hospitalier de l'Université de Montréal, Hôpital Notre-Dame, 1560 Sherbrooke Est, Montreal, Quebec, Canada.
Rationale And Objectives:
To evaluate and compare the accuracy of cross-sectional imaging using an intravascular antenna in the context of vascular morphological measurements performed during a magnetic resonance imaging (MRI)-guided vascular intervention.
Materials And Methods:
Cross-sectional imaging of a multimodality vascular phantom was performed using intravascular and surface MRI, multidetector computed tomography, and intravascular ultrasound (IVUS). Using a balanced steady-state free-precession sequence, 18 sequences parameters sets were investigated (12 for intravascular MRI and 6 for surface MRI). Vessel diameters for all images and modalities were computed using an automated vessel segmentation algorithm.
Results:
Using IVUS as a gold standard, imaging using an intravascular antenna leads to an increase in geometrical accuracy in comparison to traditional surface MRI. This level of accuracy appears to follow a significant inverse proportionality relation in respect to vessel wall signal-to-noise ratio (SNR). Taking into account the rapid decrease in SNR as a function of the distance to the intravascular antenna, these results imply that, for a given level of geometrical accuracy, faster sequences can be used for the imaging of smaller vessels.
Conclusion:
Imaging using an intravascular antenna appears as a valuable assistance to increase the accuracy of vascular morphological measurements. This increase in geometrical accuracy would be beneficial during the realization of an MRI-guided intervention, either to perform pretreatment measurements or to assess the outcome of the procedure. Acquisition parameters should be tailored to vessel size and procedural time constraints.
More Related Videos
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...