Potential preanalytical errors in whole-blood analysis: effect of syringe sample volume on blood gas, electrolyte and
Pirjo Hedberg1, Aila Majava, Kai Kiviluoma
1Laboratory, Oulu University Hospital, Department of Clinical Chemistry, University of Oulu, Finland. pirjo.hedberg@ppshp.fi
Background:
Arterial blood samples are sensitive to bias because of the physiological properties of blood. Several errors can occur in the preanalytical phase leading to incorrect diagnosis and improper treatment of patients. Collection of a blood specimen, as well as its handling and transport, belong to the key factors to affect the accuracy and good quality of clinical laboratory analysis.
Methods:
The aim of this study was to validate the effect of different sample volumes on the blood gas, electrolyte and lactate values using 3 mL Rapidlyte plastic syringes with filter cap and Rapidlab 865 blood gas analyser. Also, the stability of blood gas analyser parameters with different sample volume was studied.
Results:
No substantial change in blood gas, electrolyte and lactate parameters was found when the results of 3 mL, 1.8 mL sample volumes in the 3 mL syringes were compared. The sample volume of 1.0 mL or 1.5 mL in the 3 mL syringe is not suitable for the measurement of oxygen tension, especially when accurate results of pO(2) and arterial blood is needed for patient's diagnosis.
Conclusions:
The minimum sample volume when blood gases, electrolytes and lactate are all measured with the Rapidlab system should be 1.8 mL using 3 mL Rapidlyte plastic syringe with filtercap. According to this study <1.8 mL sample volumes can provide inaccurate results and can impose biases on measurements.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Pre-Procedural Guidelines for Assessing Blood Pressure
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...