Purpose:
In order to establish guidelines for the preparation of the Dräger Fabius GS premium anesthetic workstation for malignant hyperthermia-susceptible patients, the authors evaluated the effect of the workstation's exchangeable and autoclavable components on the washout of isoflurane.
Methods:
A Dräger Fabius GS workstation was primed with 1.5% isoflurane, and exchangeable components were replaced as follows: Group 1: no replacement (control); Group 2: autoclaved ventilator diaphragm and ventilator hose; Group 3: flushed ventilator diaphragm and ventilator hose; Group 4: autoclaved compact breathing system. The fresh gas flow (FGF) was set at 10 L . min(-1), and the concentration of isoflurane in the inspiratory limb of the circle breathing circuit was recorded every minute until an endpoint of 5.0 parts per million (ppm) was achieved, at which time the FGF was reduced to 3 L . min(-1). Six experiments were conducted in each of the four groups.
Results:
The time to achieve an isoflurane concentration of 5.0 ppm decreased in the following order: Group 1 (151 +/- 17 min) > Group 3 (137 +/- 7 min) > Group 4 (122 +/- 11 min) > Group 2 (42 +/- 6 min) (P < 0.01 vs control). Isoflurane concentration increased approximately fivefold when the FGF was reduced to 3 L . min(-1).
Conclusion:
Anesthetic washout from the Dräger Fabius GS is relatively slow. Although washout was accelerated when the Dräger Fabius GS was equipped with autoclaved components, the reduction in washout time may be less than that required for this technique to be accepted into clinical practice. A dedicated vapor-free workstation may be preferable for rapid turnover between cases.
Related Concept Videos
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
PPE Use in Healthcare Settings I: Donning