Related Experiment Video
Updated: Jun 23, 2026

Fetal Mouse Cardiovascular Imaging Using a High-frequency Ultrasound (30/45MHZ) System
Published on: May 5, 2018
Transabdominal fetal heart rate detection using NIR photopleythysmography: instrumentation and clinical results
Kok Beng Gan1, Edmond Zahedi, Mohd Alauddin Mohd Ali
1Department of Electrical, Electronic, and Systems Engineering, Faculty of Engineering and Built Environment, Universiti KebangsaanMalaysia, Bangi, Malaysia. kok_beng_gan@yahoo.com
Abstract:
In obstetrics, fetal heart rate (FHR) detection remains the standard for intrapartum assessment of fetal well-being. In this paper, a low-power (< 55 mW) optical technique is proposed for transabdominal FHR detection using near-infrared photoplesthysmography (PPG). A beam of IR-LED (890 nm) propagates through to the maternal abdomen and fetal tissues, resulting in a mixed signal detected by a low-noise detector situated at a distance of 4 cm. Low-noise amplification and 24-bit analog-to-digital converter resolution ensure minimum effect of quantization noise. After synchronous detection, the mixed signal is processed by an adaptive filter to extract the fetal signal, whereas the PPG from the mother's index finger is the reference input. A total of 24 datasets were acquired from six subjects at 37 +/- 2 gestational weeks. Results show a correlation coefficient of 0.96 (p-value < 0.001) between the proposed optical and ultrasound FHR, with a maximum error of 4%. Assessment of the effect of probe position on detection accuracy indicates that the probe should be close to fetal tissues, but not necessarily restricted to head or buttocks.

