Precision-to-tolerance capability: an important consideration in tight glucose control
Related Experiment Video
Updated: Jun 23, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Catherine Prinzing1, Sarah Rosenlund, Vicki Sukeena
1Trinity Health-Saint Alphonsus Regional Medical Center in Boise, Idaho. USA.
Objective:
To assess clinical implications of bias and variance of point-of-care glucometric measurements in cardiac surgery patients with wide variations in postoperative hematocrit.
Methods:
Point-of-care glucose measurements were compared with values from laboratory analysis of the same sample of whole blood obtained from cardiac patients early on postoperative days 1 and 2. Twenty nurses collected 89 arterial blood samples from 58 patients during a 4-month period. Bias was measured by using difference scores between paired measurements. Patients were grouped within 5% increments according to hematocrit, and analysis of variance was used to test for differences. Variation was analyzed by precision-to-tolerance analysis within 3 euglycemic tolerance ranges.
Results:
Laboratory glucose values were 62 to 224 mg/dL; point-of-care measures were 83 to 253 mg/dL. Bias was 10.85 mg/dL across all hematocrit groups. Pairs of laboratory and point-of-care glucose values differed significantly (t(174) = 10.03; P < .001). Bias increased from -2.83 mg/dL for patients with hematocrits exceeding 39% to +16.71 mg/dL for patients with hematocrits between 20% and 24%. The standard deviation of difference scores was 11.59 mg/dL overall. The difference between 5% hematocrit groups was significant (F(4) = 4.11; P = .004). Precision-to-tolerance capability ratios for specification limits of 70 to 300, 90 to 140, and 80 to 110 mg/dL were 0.30, 1.39, and 2.32, respectively.
Conclusions:
The direction of bias change between hematocrit groupings was the direction predicted in the manufacturer's information. Precision-to-tolerance measures indicated that the point-of-care equipment was not suitable for testing glucose within the planned "tighter" glycemic standards.
Related Concept Videos
Hyperglycemia
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Errors occurring during blood pressure monitoring
Several factors...
Diabetes Mellitus: Type 2 and Gestational
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.