Related Experiment Video
Updated: Jun 23, 2026

Measurement of Vibration Detection Threshold and Tactile Spatial Acuity in Human Subjects
Published on: September 1, 2016
Sub-ppm detection of vapors using piezoresistive microcantilever array sensors
Genki Yoshikawa1, Hans-Peter Lang, Terunobu Akiyama
1National Center of Competence for Research in Nanoscience (NCCR), Department of Physics, University of Basel, 4056 Basel, Switzerland. Genki.Yoshikawa@unibas.ch
Abstract:
The performance of microfabricated piezoresistive cantilever array sensors has been evaluated using various vapors of volatile organic compounds including alkanes with different chain length from 5 (n-pentane) to 14 (n-tetradecane). We demonstrate that piezoresistive microcantilever array sensors have the selectivity of discriminating individual alkanes in a homologous series as well as common volatile organic compounds according to principal component analysis. We developed a new method to evaluate the sensitivity, taking advantage of the low vapor pressures of alkanes with longer chains, such as n-dodecane, n-tridecane and n-tetradecane, under saturated vapor conditions. This method reveals sub-ppm sensitivity and the cantilever response is found to follow the mass of evaporated analytes as calculated using a quantitative model based on the Langmuir evaporation model.