Consistent relationship between automated office blood pressure recorded in different settings
Martin G Myers1, Miguel Valdivieso, Alexander Kiss
1Schulich Heart Centre, Department of Medicine, University of Toronto, Toronto, Ontario, Canada. martin.myers@sunnybrook.ca
Automated office blood pressure (BP) measurements are consistent across visits and settings. These automated readings closely align with 24-hour ambulatory BP monitoring, offering a reliable alternative to traditional methods.
Area of Science:
- Cardiology
- Hypertension Management
- Medical Device Technology
Background:
- Conventional office blood pressure (BP) measurements can be influenced by observer presence and environmental factors.
- Variability in BP readings can impact accurate hypertension diagnosis and management.
- Automated BP measurement offers a potential solution to reduce observer bias.
Purpose of the Study:
- To evaluate the consistency of automated office BP measurements during repeat visits.
- To assess the impact of different settings on automated office BP readings.
- To compare automated office BP with ambulatory BP monitoring.
Main Methods:
- Automated BP readings were taken using a BpTRU sphygmomanometer during routine hypertension specialist visits.
- Measurements were obtained before and after 24-hour ambulatory BP monitoring (ABPM).
- A third automated reading was taken at the ABPM unit visit.
Main Results:
- No significant differences were found among the three automated office BP readings.
- Automated office BP readings were comparable to mean awake ambulatory BP.
- Manual BP readings by technicians were significantly higher than mean awake ambulatory BP (P<0.001).
- High agreement was observed among automated office BP readings (ICC for systolic/diastolic BP: 0.896/0.873).
Conclusions:
- Automated office BP readings demonstrate visit-to-visit consistency.
- Automated office BP measurements are reliable regardless of the setting.
- Automated office BP readings closely approximate mean awake ambulatory BP, suggesting improved accuracy.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement