Computerized endoscopic balloon manometry to detect esophageal variceal pressure
1Department and Key Laboratory of Gastroenterology, Anhui Geriatric Institute, First Affiliated Hospital of Anhui Medical University, Hefei, Anhui, PR China.
Background And Study Aims:
Measuring the variceal pressure is important in predicting esophageal variceal bleeding. However, current noninvasive methods of measuring variceal pressure have not gained wide popularity. We have developed a computerized endoscopic balloon manometry (CEBM) system to detect variceal pressure. The aims of the study were to test the in-vitro accuracy of CEBM and to evaluate the clinical reliability and feasibility of this method.
Patients And Methods:
The CEBM system, comprising an esophageal variceal manometer and a computer, records variceal pressure and manometry images simultaneously. In the in-vitro study, variceal models were fixed inside an artificial esophagus, into which an endoscope with transparent balloon was inserted for intraluminal pressure measurement. The artificial varix was filled with water and connected to a water column to modulate the intraluminal pressure. This CEBM system was tested blindly in variceal models with different intraluminal pressures. CEBM was also used to measure variceal pressure in 23 patients with liver cirrhosis and esophageal varices, and the results were compared with the hepatic venous pressure gradient (HVPG).
Results:
In the in-vitro study, the measured intraluminal pressure correlated significantly with the actual intraluminal pressure for different diameters (R > or = 0.993, P < 0.001). Variceal pressure measurements with CEBM were technically successful in 23 patients. Regression analysis showed a good correlation between variceal pressure measured with CEBM and the HVPG (R = 0.858, P < 0.001).
Conclusions:
Our preliminary results indicate that CEBM is feasible and accurate. CEBM may become a more reliable method for detecting variceal pressure.
Related Concept Videos
Esophageal Varices-II: Clinical Features and Management
In the initial assessment, a thorough review of the patient's medical history is vital to identify risk factors such as liver disease, alcohol abuse, or...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Esophageal Varices-I: Introduction
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures III: Video Capsule Endoscopy