Pacemaker-based analysis of atrioventricular conduction and atrial tachyarrhythmias in patients with primary sinus
Martin Stockburger1, Frederike Trautmann, Aischa Nitardy
1Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Medizinische Klinik mit Schwerpunkt Kardiologie, Berlin, Germany. martin.stockburger@charite.de
Background:
Most patients with symptomatic sinus node disease (SND) receive DDDR pacemakers (PM) in order to cover SND and atrioventricular (AV) block from the outset. But the concern about adverse effects of right ventricular pacing (RVP) is increasing. So far, data on the incidence of AV block in SND are based on clinical events. The study undertakes to assess and appraise AV block and atrial tachyarrhythmias (AT) from memory and electrograms of a dual-chamber PM set to an AAIR-DDDR switch mode (AAISafeR).
Methods:
A dual-chamber PM incorporating the AAISafeR mode was implanted in 58 patients (70 +/- 10 years, 28 males) with SND, but without AV block >I. AV block and AT episodes were retrieved from the PM memory and validated from electrograms. AV block episodes were classified potentially relevant while comprising AV block III or AV block I/II during exercise.
Results:
The patients experienced a median of 90 (interquartile range 7-1,084) commutations. Possibly relevant AV block occurred in 32 patients (55%). Validation revealed high-quality PM-based categorization. The RVP prevalence was 0% (0-16%). The median AT prevalence was 0.03 (0-26) min/day. RVP was the only multivariate predictor of AT (P = 0.001).
Conclusions:
Potentially relevant AV block occurs frequently in patients with SND. Nonetheless, the RVP prevalence is kept low through the AAISafeR mode. The protection of SND patients with demand-actuated ventricular pacing appears reasonable. The AT prevalence is low in SND patients treated by the AAISafeR mode. Even low RVP proportions appear to favor AT. Prospective evaluation is needed.
Related Concept Videos
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism, and...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Electrophysiology of Normal Cardiac Rhythm