Effect of noise on T-wave alternans measurement in ambulatory ECGs using modified moving average versus spectral
Raja J Selvaraj1, Vijay S Chauhan
1Division of Cardiology, University Health Network, Toronto, Canada.
Background:
The modified moving average (MMA) and spectral method (SM) are commonly used to measure T-wave alternans (TWA), but their accuracy has not been compared in ambulatory electrocardiograms (ECGs) where TWA signal-to-noise ratio is low. Our objective was to compare the effect of noise and signal nonstationarity on the accuracy of TWA measurement using MMA versus SM when applied to synthetic and ambulatory ECGs.
Methods:
Periodic and nonperiodic noise were added to noiseless synthetic ECGs. Simulated TWA (0-20 microV) was added to synthetic ECGs and ambulatory ECG recordings. TWA was measured using SM and MMA, and the measurement error relative to added TWA was compared. An MMA ratio was used to discriminate TWA signal from noise. Signal nonstationarity was simulated by changing heart rate, TWA magnitude, and TWA phase.
Results:
With no added TWA, MMA falsely measured TWA in synthetic and ambulatory ECGs, while false measurement was not seen with SM. An MMA ratio > 1.2 eliminated false TWA detection. In the presence of low TWA magnitude (<10 microV), TWA was overestimated by MMA and underestimated by SM in proportion to the noise level. In synthetic ECGs with periodic noise and 10-microV added TWA, MMA was less accurate than SM. The effects of simulated signal nonstationarity on the TWA magnitude measured with MMA versus SM were similar using a 64-beat analysis window.
Conclusions:
In the presence of noise, MMA falsely detects or overestimates simulated TWA in ambulatory ECG recordings. In this setting, the proposed MMA ratio improves the specificity of MMA.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...