X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Noise reduction in low-dose x-ray fluoroscopy for image-guided radiation therapy
1Department of Radiation Oncology, Stanford University School of Medicine, CA 94305, USA. jingwang9@stanford.edu
Purpose:
To improve the quality of low-dose X-ray fluoroscopic images using statistics-based restoration algorithm so that the patient fluoroscopy can be performed with reduced radiation dose.
Method And Materials:
Noise in the low-dose fluoroscopy was suppressed by temporal and spatial filtering. The temporal correlation among neighboring frames was considered by the Karhunen-Loève (KL) transform (i.e., principal component analysis). After the KL transform, the selected neighboring frames of fluoroscopy were decomposed to uncorrelated and ordered principal components. For each KL component, a penalized weighted least-squares (PWLS) objective function was constructed to restore the ideal image. The penalty was chosen as anisotropic quadratic, and the penalty parameter in each KL component was inversely proportional to its corresponding eigenvalue. Smaller KL eigenvalue is associated with the KL component of lower signal-to-noise ratio (SNR), and a larger penalty parameter should be used for such KL component. The low-dose fluoroscopic images were acquired using a Varian Acuity simulator. A quality assurance phantom and an anthropomorphic chest phantom were used to evaluate the presented algorithm.
Results:
In the images restored by the proposed KL domain PWLS algorithm, noise is greatly suppressed, whereas fine structures are well preserved. Average improvement rate of SNR is 75% among selected regions of interest. Comparison studies with traditional techniques, such as the mean and median filters, show that the proposed algorithm is advantageous in terms of structure preservation.
Conclusions:
The proposed noise reduction algorithm can significantly improve the quality of low-dose X-ray fluoroscopic image and allows for dose reduction in X-ray fluoroscopy.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...